Technical Parameters
Model | TOCG-3040 |
Analytical method | UV persulfate oxidation-NDIR detection principle |
Measure parameters | TC, IC, TOC, NPOC |
TOC range | 0.001 ~ 100.0 mg/L |
Low limit | 0.001mg/L |
Accuracy | ± 5% or F.S ± 0.1% (whichever is greater) |
Analysis time | 3 ~ 8 min |
Repeatability | 1.5-10mg/L Range: ≤ 3%; |
0.1-1.5mg/L Range: ≤ 5% | |
Zero drift: | ≤ ± 2%/D |
Range drift | ≤ ± 2%/D |
Power supply | 220VAC ≤ ± 10% 50Hz ; Power: 150W |
Dimension | 56cm × 41cm × 146 cm |
Test conditions
Oxidizer | Sodium persulfate solution |
Acid reagent | Phosphoric acid solution |
Carrier gas | High-purity nitrogen (air without carbon dioxide) |
Environmental requirements | Ambient temperature: 0 ~ 40 ℃ |
Relative humidity: ≤ 85% |
Working principle
TOCG-3040 total organic carbon analyzer adopts the principle of ultraviolet persulfate oxidation-NDIR detection. One sample was decomposed into CO2 by UV and persulfate oxidation, and the other sample was decomposed into CO2 by acid reagent. CO2 is dried and purified under the action of a carrier gas, and then enters a non-dispersive infrared detector (NDIR) to be measured so as to determine the carbon concentration in the sample.
H2O + hν(185nm) →OH•+ H•
Organic matter + OH•→CO2 + H2O