First point calibration of online Ion meter
1)Put the ion electrode to the first ion buffer.
2)Press UP/DOWN key to input the correct buffer and then press ENTER to start calibration.
3)User can press ENTER to go to next when the reading is stable.
4)Display the mV on the right side.
5)If the mV is over +/-900mV or temperature is over 0.0-60.0℃ then it will display error message on the button of LCD.
Slope calibration of online Ion meter
1)Put the ion electrode to the second buffer.
2)Press UP/DOWN key to input the correct buffer and then press ENTER to start calibration.
3)User can press ENTER to go to next when the reading is stable.
4)Display the idea mV on the right side.
5)If the mV is over +/-1000mV or temperature is over 0.0-60.0℃ then it will display error message on the button of LCD.
PFG-3085 in BOQU online Ion meter can measure Fluoride ion,Chloride ion,Chloride,Potassium ion,Nitrate ion,Ammonium ion,Calcium ion,silver ion, water Hardness etc; so you can treat it as online Fluoride ion meter,online Chloride ion meter,online Potassium ion meter,online Ammonia ion meter,online Calcium ion meter,online silver ion meter,online hardness meter according to different ion.