1.Initial set of the instrument,adjust signal value in “Sys state”(system state).

2.Measurement signal 4300 (manually extract three tubes of distilled water into the digestion tank)

3.Reference signal 3000

4.High signal 1300 (volume metering device.)

5.Low signal 1300 (volume metering device.)

6.Note: If the signal value has deviation users need adjusted it at the photoelectric board at the top of the instrument